An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Metrics
Affected Vendors & Products
References
History
No history.

Status: PUBLISHED
Assigner: mitre
Published: 2020-08-26T15:14:59
Updated: 2024-08-04T13:15:20.806Z
Reserved: 2020-07-01T00:00:00
Link: CVE-2020-15486

No data.

Status : Modified
Published: 2020-08-26T16:15:12.683
Modified: 2024-11-21T05:05:37.020
Link: CVE-2020-15486

No data.