Filtered by vendor Draeger Subscriptions
Filtered by product Infinity Acute Care System Subscriptions
Total 2 CVE
CVE Vendors Products Updated CVSS v3.1
CVE-2022-4992 1 Draeger 2 Infinity Acute Care System, Standalone Infinity M540 Patient Monitor 2026-06-03 8.6 High
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors versions VG4.1.1, VG4.0.3, and lower (with VG4.2 partially affected) contain a network message handling vulnerability that allows remote attackers to inject spoofed or tampered data and cause denial-of-service conditions. Attackers can compromise network communications to modify device settings such as alarm states or alarm limits, or overwhelm the system with excessive network traffic causing the Cockpit or M540 to reboot and lose network functionality.
CVE-2019-25719 1 Draeger 2 Infinity Acute Care System, Standalone Infinity M540 Patient Monitor 2026-06-02 8.6 High
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.